This call point (S1663P) is designed for use with the
MULTIGUARD 2000 patient nurse call system (MG311000 and
MG311100). In addition to being able to generate a call by
pressing the large red button on the facia it is also possible to
reset any calls by simultaneously pressing the blue button and
the central green button/LED. An emergency recall can also be
generated at any time by simultaneously pressing the yellow
button and the central green button/LED.
The call point has an integral electronic timer which is started
whenever a call is generated. If the call has not been reset within
a predetermined period then a late attendance signal is sent to
the control equipment to re-alert staff at the nurse station. The
integral timer is adjustable over the range 30 seconds to about 2
minutes and will continue to send the signal at the periodic
intervals until the call point has been reset.
Facilities are provided for the connection of auxiliary devices
such as:-
A. an external ceiling mounted Pull-Cord with re-assurance
LED S/1600.
B. a hand held pillow push. S/1631 or other open circuit device
C. an overdoor light unit. S/1636
Details of these devices and their connection to the call point can be found overleaf.
Note: The MG311XXX controllers is a non-latching device. A pull cord, pillow push or other non-
latching call devices cannot be connected directly to the MULTIGUARD MG311XXX controllers
without an S/1663 call point. The call point provides an electronic latching circuit.

1 2 3 4 5 6 + + 1 2 3 4 + -

INPUTS

FUNCTION

INPUTS

Link F2 to +ve to disable silence function

S/1663
Complies with Council
Directive 89/336/EEC

+ve

-ve

E/Recall

L/Attend

CALL OUT

Call In

+ve

Reset In

-ve

+ve

LED

Late Attend
Timer

S/1663
Call Point

+ve

-ve

E/Recall

L/Attend

CALL OUT

Call In

+ve

Reset In

-ve

+ve

LED

Late Attend
Timer

S/1663
Call Point

+ve

-ve

E/Recall

L/Attend

CALL OUT

Call In

+ve

Reset In

-ve

+ve

LED

Late Attend
Timer

S/1663
Call Point

+ve

-ve

E/Recall

L/Attend

CALL OUT

Call In

+ve

Reset In

-ve

+ve

LED

Late Attend
Timer

S/1663
Call Point

Pillow push or Pull Cord
Input. Normally Open, Close
to Call

Output for Re-assurance LED
on Pull Cord or Overdoor Light

See overleaf for connection
details

MG311000
& MG311100

+ve

-ve

E/Recall

L/Attend

CALL OUT

Call In

+ve

Reset In

-ve

+ve

LED

Late Attend
Timer

S/1663
Call Point

CALL

RESET RE-CALL

MULTIGUARD CALL/RESET
POINT S1663P

The call point is
designed to fit into
a standard single
gang pattress box
or flush mount
accessory box.
A box of 25mm or
greater depth is
recommended.
Eg. S/1680 or
TE55

A 10 core cable can be used for
each six zones. It is possible to
connect a number of call points
into one zone of the annunciator

To additional

call points

To additional

call points