Differential output measurements – Kaman SMU9200 User Manual

Page 15

Advertising
background image

SMU-9000 User Manual

12

••••

Part III: Calibration

from the target, the second sensor would therefore be positioned at
minimum displacement respectively.

Differential Output Measurements

In a differential calibration, each channel of the SMU-9000 is
calibrated individually. After calibrating each channel, a
“balanced to ground” output is taken from the two the positive
labeled output wires in order to display the differential output
voltage. To accomplish this, connect the positive output wires to a
voltmeter as shown in Figure 10.

DVM

P/O CABLE

WHITE CH A +

BLUE CH B -

ORANGE CH B +

GREEN CH A -

+ INPUT

NC

NC

- INPUT

Figure 10. Differential output wire connections

This “balanced ground” connection allows the output of one
channel to be subtracted from the other, producing a differential
voltage reading at the voltmeter. For example, the differential
output voltage measurements of a system calibrated using this
method may look similar to the following:

Table 2. Differential Output Readings

Channel A Output

Channel B Output

Ch. A - Ch. B

0 V

+10 V

-10 V

+6.3 V

+6.3 V

0 V

+10 V

0 V

+10 V