Kaman Extreme Environment Sensors User Manual

Page 18

Advertising
background image

High Temperature Displacement System

Troubleshooting

Troubleshooting the system is limited primarily to isolating the malfunction between
the various components.

The sensor can be checked for coil integrity as follows.

A.

Disconnect the solid sheathed cable from the flexible cable.


B.

Using and ohm-meter that has a battery voltage of 1.5 volts or less

check continuity form each individual pin of the connector to the
sheath. These values should be approximately 4.5 ohms at room

temperature.


Anomalous readings could indicate failure at the cable connector, the cable, or within

the sensor. Time-Domain-Reflectometer techniques can be used to isolate the failure
region. Generally only the connector is readily repairable.

The flexible cable can likewise be checked in a similar manner by attaching the
sensor and taking corresponding resistance measurements.

Minor changes in sensitivity are covered in the sections pertaining to installation,
calibration, and adjustment. Some changes may be attributable to changes in

grounding modified mounting or installation, and high resistance contacts. The first
two conditions may require readjustment of the electronics to maintain desired

sensitivity and linearity. The latter will require contact cleaning as described in the

maintenance section.






-17-