Kaman ThreadChecker User Manual

Page 9

Advertising
background image


Copyright © 2008

Kaman Aerospace Corporation

PART NO: 860510-001A

Kaman Precision Products

Last Revised: 6/8/11

Measuring & Memory Systems

217 Smith Street

Middletown, CT 06457

www.kamansensors.com or 860-632-4442

Page 9 of 13


Push Button Switch Operation and User Configuration

Switch Operation

Result

Holding the button down for longer than
one second

The status light starts blinking. On release it
acquires the air readings for all frequencies.
The next step is to put the sensor in a threaded
hole and press the button again. On release
the system will acquire a set of calibration data
for all frequencies for the threaded hole. The
last step is to put the sensor in a non-threaded
hole and press the button again. On release
the system will acquire data for all frequencies
for the non-threaded hole, optimize the sensor
frequency and modality, and set the limits.

Holding switch down during power up for
less than 10 seconds

Toggles between window comparator and level
comparator

Holding switch down during power up for
more than 10 seconds

Status LED will turn yellow and when the
switch is released it will toggle the Dwell
Enable (i.e. whether or not you are using the
derivative to check the limits)

Holding switch down for longer than 10
seconds (until the status light stops
blinking)

Limits are inverted

Advertising