Calibration tab, Samples, 4 calibration tab – PNI CompassPoint Prime User Manual

Page 22: 1 samples

Advertising
background image

PNI Sensor Corporation

DOC# 1014177 r08

Prime User Manual

– July 2011

Page 16

5.4

Calibration Tab

5.4.1

Samples

Before proceeding, refer to Section 6.2 for the recommended calibration procedure

corresponding to the calibration method selected on the Configuration tab.

Clicking the <Start> button begins the calibration process.

If “Automatic Sampling” is not checked on the Configuration tab, it is necessary to click
the <Take Sample> button to take a calibration sample point. This should be repeated

until the total number of samples (as set on the Configuration tab) is taken, changing the

orientation of the module between samples as discussed in Section 6.2.

If “Automatic Sampling” is checked, the module will need to be held steady for a short
time and then a sample automatically will be taken. Once the window indicates the next

number, the module’s orientation should be changed and held steady for the next sample.

Once the pre-set number of samples has been taken (as set on the Configuration tab) the

calibration is complete.

Advertising