ROTRONIC HC2-Ix25 flush mount probe User Manual

Hc2-ix25

Advertising
background image

12.0980.0001

ROTRONIC AG, CH-8303 Bassersdorf
Tel. +41 44 838 11 44, www.rotronic.com
ROTRONIC Messgeräte GmbH, D-76275 Ettlingen
Tel. +49 7243 383 250, Fax +49 7243 383 260, www.rotronic.de
ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg
Tél. +33 1 60 95 07 10, www.rotronic.fr
ROTRONIC Italia srl

,

I- 20157 Milano

Tel. +39 2 39 00 71 90, Fax (+39) 02 33 27 62 99, www.rotronic.it

ROTRONIC Instruments (UK) Ltd, West Sussex, RH10 9EE , UK
Phone +44 1293 571000, www.rotronic.co.uk
ROTRONIC Instrument Corp, NY 11788, USA
Phone +1 631 427-3898, www.rotronic-usa.com
ROTRONIC South East Asia Pte Ltd, Singapore 339156
Phone +65 6294 6065, www.rotronic.com.sg
ROTRONIC Shanghai Rep. Office, Shanghai 200233, China
Phone +86 40 08162018, www.rotronic.cn

R O T R O N I C H

ygro

C

lip

®

HC2-Ix25

Mit dem ROTRONIC HygroClip HC2-Ix25 haben Sie ein Spitzenprodukt erwor-

ben, das den gängigen Normen entspricht. Bei Beachtung der unten aufge-

führten Bedienungshinweise funktioniert der HygroClip über Jahre perfekt.

Einsatzbereich Feuchte:

0...99 %rF

Einsatzbereich Temperatur:

HC2-IP25/HC2-IS25/HC2-IT25: -40...85 °C
Standard Messbereiche:

Alle Parameter sind frei skalierbar

Feuchte:

0...100 %rF = 0...1 V

Temperatur:

-40…60 °C = 0...1 V

Tau-/Frostpunkt:

Einstellbar via PC&HW4 Software

Technische Daten:

Versorgung:

3,2...5 VDC ± 0 % / ca. 4 mA

Justiert bei 3.3 VDC. Bie anderen

Versorgungsspannungen muss

der Fühler neu justiert werden

Ausgangsbürde:

>10 k

Kabellänge:

Max. 5 m ohne Booster

Max. 100 m mit Booster Art. Nr.: AC3003

Schutzart:

IP65 (ausser Sensorbereich)

Ausgänge:

Analog: 2 x 0…1 VDC, UART Schnittstelle

Genauigkeit bei 23 °C:

0...90 %rF 1.5 %rF, 0.2 K

90...99 %rF 2.0 %rF, 0.2 K

Montage

Die Fühler sind dafür vorgesehen in eine Schaltwand oder Paneele

eingebaut zu werden.

Maximale Wandstärke:

9 mm

Einbauöffnung:

ø 25 mm (25…27mm)

Die Schutzkappe dient zum Transportschutz und zur Abdeckung des

Messraumes im Falle einer Reinigung oder Desinfektionsmassnahme der

Schaltwand oder Paneele. Für die Inbetriebnahme und Feuchtemessung

des Fühlers ist die Schutzkappe zu entfernen.

R O T R O N I C H

ygro

C

lip

®

HC2-Ix25

Your ROTRONIC HygroClip HC2-Ix25 is a leading-edge product that meets

all established standards. It will work perfectly for many years if you follow

the operating guidelines below.

Operating range humidity:

0...99 %rh

Operating range temperature:

HC2-IP25/HC2-IS25/HC2-IT25: -40...85 °C
Standard measuring ranges:

All parameters are freely scaleable

Humidity:

0...100 %rh = 0...1 V

Temperature:

-40…60 °C = 0...1 V

Dew-/frost point:

Selectable by PC using HW4 software

Technical data:

Supply:

3.2...5 VDC ± 0 % / approx. 4 mA

Adjustet at 3.3 VDC. The probe needs to

be adjusted when the probe is powered

on a different voltage than 3.3 VDC.

Load:

>10 k

Cable length:

Max. 5 m without booster

Max. 100 m with booster item no.: AC3003

Protection rating:

IP65 (except sensor area)

Outputs:

Analog: 2 x 0…1 VDC, UART interface

Precision at 23 °C:

0...90 %rh 1.5 %rh, 0.2 K

90...99 %rh 2.0 %rh, 0.2 K

Mounting

The probes are intended for being built into walls or panels.

Maximum wall thickness:

9 mm

Cutout:

ø 25 mm (25…27mm)

The protection cap is used for transport protection and coverage of the

probe in case of cleaning. For commissioning and measurement the

protection cap is removed.

Anschluss

Setzen Sie den HygroClip auf den Gegenstecker des Messumformers oder

des Anschlusskabels. Achten Sie darauf, dass die Sperrnocken richtig

ausgerichtet sind. Ziehen Sie die Rändelmutter von Hand fest.
Justierung / Kalibrierung

Die Fühler sind ab Werk justiert. Wir empfehlen je nach Einsatz eine

Kalibrierung nach ca. einem Jahr. Dies kann vor Ort mit ROTRONIC SCS

Feuchte-Standards durch Sie erfolgen. Die Handhabung entnehmen Sie

der den Feuchtestandards beiliegenden Anleitung. Betreffend Justierung

kontaktieren Sie ROTRONIC oder unsere Vertretung.

Weitere Information unter www.rotronic.com

Connection

Plug the HygroClip on the mating connector of the transmitter or connec-

tion cable. Make sure that the catches are aligned correctly. Tighten the

knurled nut by hand.
Adjustment / Calibration

The probes are adjusted in our factory before delivery. We recommend

calibration after about one year depending on use. You can do this yourself

on-site with the ROTRONIC SCS humidity standards. Please refer to the

instructions included with the humidity standards. Contact ROTRONIC or

our representatives concerning adjustment.

Further information can be obtained at www.rotronic.com

Electrical connections:

Elektrische Anschlüsse:

1 green V +

2 gray GND (digital and power)

3 red

RXD (UART)

4 blue TXD (UART)

5 white Analog signal, humidity

(0...100 %rh = 0...1 V)

6 brown Analog signal, temperature

(-40...60 °C = 0...1 V)

7 yellow AGND (analog ground)

1 grün V +

2 grau GND (Digital und Versorgung)

3 rot

RXD (UART)

4 blau TXD (UART)

5 weiss Analogsignal Feuchte

(0...100 %rF = 0...1 V)

6 braun Analogsignal Temperatur

(-40...60 °C = 0...1 V)

7 gelb AGND (analog ground)