ROTRONIC HF56 User Manual

ROTRONIC Accessories for electrical

Advertising
background image

12

.0

97

3.0

00

2

ROTRONIC AG, CH-8303 Bassersdorf
Tel. +41 44 838 11 44, www.rotronic.com
ROTRONIC Messgeräte GmbH, D-76275 Ettlingen
Tel. +49 7243 383 250, Fax +49 7243 383 260, www.rotronic.de
ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg
Tél. +33 1 60 95 07 10, www.rotronic.fr
ROTRONIC Italia srl

,

I- 20157 Milano

Tel. +39 2 39 00 71 90, Fax (+39) 02 33 27 62 99, www.rotronic.it
ROTRONIC Instruments (UK) Ltd, Crompton Fields,
Phone +44 1293 571000, www.rotronic.co.uk
ROTRONIC Instrument Corp, NY 11788, USA
Phone +1 631 427-3898, www.rotronic-usa.com
ROTRONIC South East Asia Pte Ltd, Singapore 339156
Phone +65 6294 6065, www.rotronic.com.sg
ROTRONIC Shanghai Rep. Office, Shanghai 200233, China
Phone +86 40 08162018, www.rotronic.cn

A

Digital transmitter for humidity & temperature
Duct & Wall and Cable version

Congratulations on your purchase of the new state-of-the-art HygroFlex5-series transmitter.
Please read these short instructions carefully before installing the device

General description

The HygroFlex5-series devices are universal transmitters for transmission of humidity and tem-
perature measurements. Compatible with all interchangeable HC2 probes. These short instructions
are limited to a description of the main functions and installation of the device.
The detailed instruction manual can be found on the internet at:

www.rotronic-humidity.com

Dimensions / Connections

Wall version (Type W)

Mechanical installation

General recommendations
Relative humidity is extremely temperature-dependent. In order to measure it exactly, the probe and
sensors must be set exactly on the temperature level of the environment that is to be measured.
The installation site can therefore have a significant influence on the performance of the device.
Follow the guidelines below to ensure optimum performance:

a) Select a representative installation site: Install the probe at a point where the humidity, tem-

perature and pressure conditions are representative for the environment that is to be measured.

b) Make sure there is sufficient air movement around the probe: An air flow of at least 1 metre/

second accelerates and facilitates adjustment of the probe to changing temperatures.

c) Avoid:
1. Probe too close to heating elements, cooling coils, cold or hot walls, direct sunlight, etc.
2. Probe too close to steam, injectors, humidifiers or direct precipitation.
3. Unstable pressure conditions with high air turbulence.
d) Insert the probe as far as possible into the environment that is to be measured.
e) Avoid accumulation of condensation at the contact wires of the sensor. Install the probe so

that the tip points down. If that is not possible, install it in horizontal position.

AC5005 Mounting flange
for 15 mm probe

Terminal K6-1: Earth is usually connected to GND. If this is not wanted, a pad on the PCB (

B2) must

be removed.

Terminals K3 (RS-485): Terminals K3-1 and K3-2 can be used to feed the device (multi-point
connection). Several RS-485 devices can be operated with a strong 15 VDC power supply unit.
In this case the supply voltage at K6-1 to K6-3 is not used.

Warning: Make sure that all settings have been made correctly before integrating and connecting
the transmitters in the network.

Programming

The basic settings of the devices are made in the factory according to your order. The transmit-
ters are adjusted in the factory and therefore do not need to be checked and readjusted during
installation. The devices can be started immediately after installation.

Display / Keypad
The LC display has a backlight which can be set to be on all the time or whenever a key is pressed.
The backlight can also be disabled.

Press the Enter button to switch between 2 and 3-line display.

The display can also be configured to show a trend indicator on each line:

▲: increasing value

▼: decreasing value
In the event of an alarm “Sensor Alarm” appears on the bottom of the display.

Unity system can be switched between Metric and English.

3 / 4 wire circuit / HF56x

Button MENU

open / close menu

Button ENTER
select menu point

Menu navigation
Buttons + / - change value
increase/decrease

Note: Unauthorised use of the menu can be prevented by locking the setting “Display Menu”
(using the HW4 software > Device Manager > Display).

Sources of error

Measured values can be influenced by the following factors:

Temperature errors :
Adaptation time too short, cold outside wall, heating elements, sunlight, etc.
Humidity errors:
Steam, water spray, dripping water or condensation at the sensor, etc. Repeatability and long term
stability are, however, not influenced by these factors even if the probe is exposed to high humidity
or saturation with steam (condensation) over a longer period of time.
Soiling:
By dust in the air. The choice of probe filter depends on the amount of soiling at the measuring
point. The filter must be cleaned or replaced periodically.

Scaling / Adjustment / Firmware update
The following settings can be made with the help of the HW4 software and either the service cable
AC3006 or AC3009:
• new scaling of the outputs
• adjustment
• firmware update

You can find a detailed description in the manual that you can download from our web site at
www.rotronic-humidity.com

Periodic calibration of the probe / transmitter
Both the Pt 100 RTD temperature sensor and the corresponding electronics are very stable and do
not normally need to be changed or calibrated after factory calibration. The long term stability of the
ROTRONIC Hygromer

®

humidity probes is typically better than 1 %rh per year. For maximum accuracy

we recommend calibration of the probe about every six to 12 months. More frequent calibration
can be necessary in applications where the sensor is exposed to pollutants. The calibration can
be performed by the user himself on site or in the laboratory / workshop. For routine calibrations
the probe should be checked at one or two points.

The electronics of the transmitter do not normally require calibration in the field. They can be
checked easily with the help of the probe simulator in the HW4 software package. The electronics
cannot be repaired in the field and should be returned to the manufacturer in the case of problems.
For details on calibration, please see the full version of the instruction manual, which you can
download from the internet.

Technical data (measurement)

Humidity:

0...100 %rh

Temperature:

Probe dependent: -50...100°C

Accuracy:

Probe dependent: ±0.8 %rh, ± 0.1 K @ 23°C

Protection:

IP65 except for models with USB and Ethernet interface

Outputs:

Current or voltage signals, digital output depending on order code,

UART service interface


Technical data
(operation)
Temperature:

–40...60 °C / Models with display –10...60 °C

Humidity:

0...100 %rh, non-condensing

Technical data probe
Depending on type

Current output

Voltage output

SHoRT iNSTRUCTioN MANUAl

Mounting the wall version

Alignment
Mount the transmitter so that the probe points down.

Mounting variant 1
Drill the necessary holes using the drill template drawn on the
packaging. Then insert the plugs delivered with the device and
mount the transmitter with the screws.

Mounting variant 2
If there is a TS35 DIN top-hat rail available, the transmitter can be
clipped on to the top-hat rail directly with the help of the mounting
kit AC5002 (available as optional extra). For this, the DIN holders
(a kit has two holders and eight screws) are screwed directly on
to the predrilled holes in the transmitter.

Electrical installation

Power supply
3-wire galvanic separated with analogue outputs: 85 to 240 VAC.
When both outputs are connected, the maximum current consmption is 50 mA.

Supply voltage / Technology
Type

Supply voltage V+

load

output

3 / 4-wire
HF561

85…240 VAC

Max. 500

Ω

0...20 mA

HF562

85…240 VAC

Max. 500

Ω

4...20 mA

HF563

85…240 VAC

Min. 1000

Ω

0...1 V

HF564

85…240 VAC

Min. 1000

Ω

0...5 V

HF565

85…240 VAC

Min. 1000

Ω

0...10 V

Caution:

Wrong supply voltages and excessively high loading of the outputs can

damage the transmitter.

Duct version (Type D)

Terminal Description
K2-1: OUT1

Analogue humidity output +

K2-2: OUT2

Analogue temperature output +

K2-3: GND

Ground (tied with other GND)

K2-4: GND

Ground (tied with other GND)

K3-1: PWR

not used

K3-2: GND

Ground (tied with other GND)

K3-3: D+

RS-485 Bi-directional TX– / RX –

K3-4: D–

RS-485 Bi-directional TX+ / RX +

K6-1:

Protective ground

K6-2: + DC+ / AC L

Power supply phase, 85…240 VAC

K6-3: – DC– / AC N

Power supply neutral, 85…240 VAC

Mounting the duct version (Type D)
To avoid measurement errors, at least 200 mm of the probe should be inserted into the environ-
ment that is to be measured.
If necessary, use the mounting flange AC5005 to install the probe.

Service interface
(Mini-USB)

Digital connection

Terminal configuration / Connection diagrams
The type is defined using the table Supply voltage / Technology to then use the following con-
nection diagrams:

Signal

Power

Advertising