Compensated Temperature Range

Model

Electrical Connector

Temperature

540 / 542

10-6 Bayonet Connector

-65°F to +250°F (-54°C to +120°C)

541

10-6 Bayonet Connector

-65°F to +390°F (-54°C to +200°C)

540 / 542

Weatherproof Cable

-65°F to +250°F (-54°C to +120°C)

540 / 542

Molded Immersible Cable

-65°F to +250°F (-54°C to +120°C)

541

10-5 Screw Lock Connector

-65°F to +390°F (-54°C to +200°C)

Operating Temperature Range

Model

Electrical Connector

Temperature

540 / 542

10-6 Bayonet Connector

-65°F to +275°F (-54°C to +135°C)

541

10-6 Bayonet Connector

-65°F to +385°F (-54°C to +195°C)

540 / 542

Weatherproof Cable

-65°F to +250°F (-54°C to +120°C)

540 / 542

Molded Immersible Cable

-4°F to 122°F (-20°C to +50°C)

541

10-5 Screw Lock Connector

-65°F to +450°F (-54°C to +230°C)

OPERATIONAL LIFE:

Limited to 100M cycles to maximum allowable pressure.

CALIBRATION

Transducers are calibrated to the datum requested at time of order; this can be identified as follows:-

Code G - gauge datum vented to atmosphere via the electrical connection

Code A - absolute datum

Code S - sealed reference; reference side of the instrument is sealed and the output electrically adjusted
to zero with 1013 mb applied to pressure port

Code U - uni-directional differential

Code B - bi-directional differential

CHARACTERISTICS

Millivolt Output Transducers: All millivolt (mV) output transducers should be used with a
stabilized d.c. excitation supply. Output errors directly proportional to supply instability.

Specifications are given with 10V d.c. excitation. However, a lower supply may be used with a
corresponding decrease in output.

There are no adjustments for zero and span on the transducer. If local provision is required, the circuit
diagram below may be employed. 12V d.c. stabilized supply is used instead of standard 10V to allow
span adjustment to be obtained.