3.0 ELECTRICAL INSTALLATION

Wiring is through factory installed PG-9 Strain Reliefs. Wiring terminations are
identified below the terminals (Diagram 1).

NOTE: Prep Transducer and Data Acquisition Wires Prior to Wiring to the 299: Strip transducer
cable back to expose wires and capillary vent tube. Tube should remain free and not crimped or
pinched. Strip data acquisition cable to expose wires. CAUTION: Overtightening of the PG9 Strain
Reliefs may pinch tube and cables.

3.1 Current Unit

The Model 299 (current input) - Reverse polarity protected when used with a
Setra pressure transducer.

The electrical connections are as follows:

3.2 Voltage Unit

The Model 299 (voltage input)

electrical connections are as follows:

Model 299

Setra Termination Enclosure

(Voltage Input)

Pressure

Transducer

+

+

_

out

out

GRND

_

Data

Acquisition

System

+

+

_

out

GRND

out

_

exc.

exc.

_

+

_

+

+

exc.

exc. _

_

+

GRND

GRND

GRND

Model 299

Setra Termination Enclosure

(Current Input)

Pressure

Transducer

+

+

_

exc.

exc.

GRND

_

Data

Acquisition

System

+

+

_

exc.

GRND

exc.

_

GRND

-

Diagram 1

Diagram 2

Diagram 3

-

+

-

+

GRND

Out

Out

Exc

Exc

GRND

Out

Out

+

Exc

Exc

-

+