0 model 223 performance specifications – Setra System Model 223 User Manual

Page 6

Advertising
background image

BAR COMPOUND

CURRENT

RANGE

OUTPUT (mA)

-1 to 1.7

9.92

-1 to 3.4

7.63

-1 to 7

6.00

-1 to 17

4.88

-1 to 35

4.44

-1 to 70

4.22

-1 to 200

4.07

4.3 Current Output Zero Adjustment

While monitoring the current output , and with the pressure port open to
atmosphere or with zero pressure applied, the zero may be adjusted by turning
the zero potentiometer screw. The factory setting is 4mA (

±0.08 mA). For

absolute pressure ranges, the zero should be adjusted with a full vacuum
applied. For compound pressure ranges, the zero should be adjusted with
the unit at atmospheric pressure and the output adjusted as noted in
Diagram 5 below.

Note: When it is not possible to vent input pressure for zero adjustment (ie., hazardous
gasses), zero may be adjusted by pulling a vacuum and setting the unit to 4.0 mA (

±0.08 mA).

This method may result in a small zero setting error due to variations in barometric pressure.

4.4 Current Output Span Adjustment

Span or full scale output adjustments should only be performed by using an
accurate pressure standard (electronic manometer, digital pressure gauge, etc.)
with at least comparable accuracy to the 223 transducer (

±0.25% FS). With full

range pressure applied to the pressure port, the span may be adjusted by
turning the span potentiometer screw. The span (full scale) output is factory
set to within

±0.08mA.

5.0 MODEL 223 PERFORMANCE SPECIFICATIONS

Accuracy RSS

*

(at constant temperature.)

±0.25% FS

Non-Linearity, BFSL

±0.15% FS

Hysteresis

0.20% FS

Non-Repeatability

0.02% FS

*RSS of Non-Linearity, Non-Repeatability and Hysteresis.
Thermal Effects
Compensated Range

°F(°C)

+15 to +150 (-9 to +65)

Zero Shift %FS/100

°F (50°C)

±2.0 (±1.8)

Span Shift %FS/100

°F (50°C)

±2.0 (±1.8)

Warm-up Shift

<

±0.1% FS total

Environmental
Temperature
Operating**

°F(°C)

-40 to +185 (-40 to 85)

Storage

°F(°C)

-40 to +185 (-40 to 85)

Current unit ordered with Option N1:
Operating**

°F (°C)

-22 to +176 (-30 to +80)

Storage

°F (°C)

-22 to +176 (-30 to +80)

** Operating temperature limits of the electronics only. Pressure media temperature may be

considerably higher or lower.

NOMINAL “INSTALLED” OUTPUT AT ATMOSPHERIC PRESSURE

PSI COMPOUND

CURRENT

RANGE

OUTPUT (mA)

-14.7 to 25

9.92

-14.7 to 50

7.63

-14.7 to 100

6.05

-14.7 to 250

4.89

-14.7 to 500

4.46

-14.7 to 1000

4.23

-14.7 to 3000

4.08

Diagram 5

6

Advertising