2

Setra Model 224

Ultra High Purity Flow-Through Pressure Transducer

1.0 GENERAL INFORMATION

Every Model 224 has been tested and calibrated before shipment. Model 224
performance specifications are listed on Page 6 of this Guide.

Setra Systems 224 pressure transducers sense gauge, compound or absolute
pressure and convert this pressure to a proportional high level analog output.
Two output versions are offered: A voltage output of 5 VDC FSO (Full Scale
Output) and 10 VDC FSO ), and a current output of 4 to 20 mA.

Note: These instructions are also available on our website at www.setra.com.

1.1

EMC Certification

This product complies with EN61326 Electrical Equipment for Measurement,
Control and Laboratory use – EMC Requirements for Minimum Requirements
and Industrial Locations. Special caution should be taken to meet Standard
EN61000-4-5: 1995 Surge Immunity if any of the following conditions apply to
the installation: The product is installed outside; all or any part of the cable is
exposed to the outside; the cable is greater than 30 meters in length. In order
to meet the Surge Immunity requirements, the following conditions must be
followed during installation:

1. Shielded cable must be used, and the shield must be tied to earth ground (not power supply

ground) on at least one end of the cable shield/drain wire. The shield must be maintained all the

way from sensor to the power supply.

2. If unshielded cable is used, an earth grounded metal conduit fitting can be used to replace the

shielded cable.

3. For a sensor with a metal body or enclosure, the body/enclosure must be grounded to earth. If a

protective metal housing is used, the metal housing should be grounded to earth

4. If a protective plastic housing is used, the housing must be able to withstand at least

2 KV from the housing to earth ground, without damaging the circuit.

2.0 MECHANICAL INSTALLATION

2.1 Media Compatibility

Model 224 transducers are designed to be used with any gas or liquid compat-
ible with 316L Stainless Steel. Never submerge the transducers in any liquid.

2.2 Environment

The operating temperature limits of the 224 are as follows:

Operating Temperature Range

°F (°C)

-40 to +185 (-40 to +85)

Compensated Temperature Range

°F (°C)

+15 to +150 (-9 to +65)

Current unit ordered with Option N1:
Operating Temperature Range

°F (°C)

-22 to +176 (-30 to +80)

Compensated Temperature Range

°F (°C)

+15 to +150 (-9 to +65)

2.3 Pressure Fittings

Mounting - Model 224 pressure transducers can be installed with Face Seal
Fittings or can be butt welded into place.
When installing units with Face Seal Fittings:

a.) Align piping system to transducer connections
b.) Hand tighten nuts
c.) Torque nuts by placing wrenches on nuts only (or on nut and wrench flats
of transducer body for units with fixed male face seals) Never hold the unit
by the electronics housing during installation.

NOTE: Welding purge gases are very hot! It is important to flow hot purge gases away from the
transducer. Disconnect all electrical connections to transducer when welding.