3.3 Remote Sensors Installation

Cable Version (P/N: 231GRSX4MXXXX)

For cable version remote sensors, use the following instructions:

1. Install the remote sensors into the system plumbing (use 3/4” hex to tighten”)

Important:

Each remote sensor is labeled “HIGH PRESSURE SENSOR” or “LOW PRESSURE SENSOR”, respectively. Match up

the high pressure sensor to the high “HI (+)” pressure electrical terminals, and the low “LO (-)” pressure

sensor to the low pressure electrical terminals. Otherwise, the performance may not meet the product

linearity specifications.

2. Attach the cable provided to the 3-Pin Packard connector on the remote sensor.

3. Feed the cable through the PG fitting on the bottom of the unit and tighten the fitting.

4. Remove the terminal blocks from the circuit board and use the following chart:

Important: Prior to putting the unit into service, complete configuration setup in Section 4. After configuration is

complete, press the “Zero” button.”

IN

MM

3.00

76.0

1.46

37.1

0.325

8.3

1.62

41.2

Chart 2. Electrical wiring color chart