EA003 OPTICAL TACHOMETER CALIBRATION ADAPTER

OPERATION MANUAL

Transmille Ltd.

Page 7


2.

The optical tachometer should be pointed at the LED on the adapter with no more

than 20° angle to ensure consistent detection by the tachometer. Distance from the

LED should also be restricted to within 5cm, again, for consistent detection.

Optical tachometer calibration times can be significantly reduced by using the ProCal

calibration software available from Transmille which allows a pre-defined sequence of tests

(known as a procedure) to be set up. This allows the computer to automatically step

through these tests, control the calibrator, set the correct outputs and record the amount of

deviation in relation to the instrument’s specifications.