8500 LOW THERMAL SCANNER
OPERATION MANUAL

Transmille Ltd.

Page 11

Remote Channel Select


The 8500 Scanner can be fully controlled via the RS232 interface. The interface

uses the standard 9 pin PC connector and a standard serial lead. The scanner can

be sent individual commands directly from the Windows HYPER TERMINAL

program, any basic or high level program or from the ProCal Calibration System

To allow communication from PC based programs, these programs must be

configured to the following settings :

BAUD RATE :

9600

PARITY : NONE

DATA BITS :

8

STOP BITS :

1

Once connected, using the PC the channel number should be sent to the scanner to
set the output. The command MUST be terminated with a carriage return
(ASCII character 13), for example :

COMMAND DESCRIPTION
1

<CR>

Set Channel 1

2

<CR>

Set Channel 2

3

<CR>

Set Channel 3

4

<CR>

Set Channel 4

5

<CR>

Set Channel 5

6

<CR>

Set Channel 6

7

<CR>

Set Channel 7

8

<CR>

Set Channel 8

9

<CR>

Set Channel 9

10

<CR>

Set Channel 10


Where <CR> is a carriage return (ASCII 13)