TEMPERATuRE CALIBRATION

A process to set temperature calibration to an ice

point is included to compensate for differences in

temperature sensors, or to reset the unit to proper

operation. Please follow these steps to perform a

temperature set;

1. Connect temperature probe

2. Press “Pressure TEMP” to display probe

temperature (verify that “TEMP” is shown on

the LCD.

3. Place the temperature probe sensing area in a

mixture of crushed ice and water. (Distilled

water will provide the most accurate source)

4. Stir the ice water mixture with the temperature

probe and allow the reading to stabilize

5. Press and hold “Temp SH/SC” button for four

seconds

This will set the ice point to 32°F. The meter will

compensate for changes up to 10°F.