KANE900 Plus Operators Manual

Page 11

4.6 Regular Checks During Sampling


Care must be taken at all times not to exceed the analysers operating specifications, in
particular ensure the following :-

• Do not exceed the maximum temperature of the flue probe.

• The analyser internal temperature does not exceed normal operating range,

typically 0-50

°C.


• DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE.

• The water trap is vertical at all times. Water condenses in the probe line and can

quickly fill the water trap when the probe is moved. Take care and watch the water
trap closely.


• The in-line particle filter is clean and does not become blocked.

4.7 Normal Shutdown Sequence

DO THIS EVERY TIME YOU USE THE ANALYSER

Remove the probe from the flue - TAKE CARE ! THE PROBE WILL BE HOT -
and allow it to cool naturally. Do not immerse the probe in water as this will be drawn
into the analyser and damage the pump and sensors.

Once the probe is removed from the flue press

and the analyser will count

down from 30 to switch off.

OFF 30

MENU TO ESCAPE

If you have not finished but press

by mistake, you can press

to

return to normal operation and not switch OFF.