Specifications – UEi Test Instruments THL2 User Manual

Page 10

Advertising
background image

9

Specifications

Relative Humidity Overall range

0 to 100%

Accuracy (0 to 20 and 80 to 100%)

±5.0%

Accuracy (20 to 40 and 60 to 80%)

±3.5%

Accuracy (40 to 60%)

3.0%

Temperature

Overall range

-40˚ to 158˚F

(-40˚ to 70˚C)

Accuracy (-40˚ to +14 and 104˚ to 158˚F) ±3.6˚F

Accuracy (+14˚ to +104˚F)

±1.8˚F

Accuracy (-40˚ to -10˚ and +40˚ to +70˚C) ±2˚C

Accuracy (-10˚ to +40˚C)

±1˚C

Dew Point

Overall range

-40˚ to 158˚F

Temperature

(-40˚ to 70˚C)

Accuracy (25˚C, 40 to 100% RH)

±4.0˚F (2.0˚C)

Logging Rate

Selectable sampling interval: From 2 seconds up to

24 hours.

Operating

-31˚ to 175˚F (-35˚ to 80˚C)

Temperature
Battery Type

3.6V lithium (1/2AA) (SAFT LS14250, Tadiran TL-5101

or equivalent)

Battery Life

1 year (typ.) depending on logging rate, ambient

temperature and use of alarm LEDs.

Dimensions

101 x 25 x 23 mm (4 x 1 x 9”)

Weight

172 g (6 oz)