2 description of the functions – A&D TM-2655 User Manual

Page 23

Advertising
background image

15

7-2 Description of the Functions

Table 5 Functions

No. Function SYSTOLIC

display

DIASTOLIC

display

Description

F01

f01 off

Not available.

off

No printing

1

High-speed printing

2

3-line printing

3

3-line printing

4

Graph printing

F02

f02

5

Table printing

off

No bit pattern printing

F03

Printer setting

(See “7-3
Printing
Samples” for
what each
method printing
looks like.)

TM-2655P only

f03

on

Bit pattern printing

999

The results remain in the display unless
another operation is performed.

5

Displays the results for 5 seconds.

10

Displays the results for 10 seconds.

F04 Displaying

time

f04

20

Displays the results for 20 seconds.

1

Terminals 1 and 2 to a PC

2

Terminal 1 to a PC
Terminal 2 to a card reader

F05 External

connection

Ff05

3

Terminal 1 to a scale
Terminal 2 to a PC

au

Automatic pressurization

160

Pressurizes the cuff up to 160 mmHg.

180

Pressurizes the cuff up to 180 mmHg.

F06

Pressurization

Ff06

200

Pressurizes the cuff up to 200 mmHg.

120

1200 bps

240

2400 bps

480

4800 bps

F07

Channel 1 baud
rate

Ff07

960

9600 bps

120

1200 bps

240

2400 bps

480

4800 bps

F08

Channel 2 baud
rate

Ff08

960

9600 bps

F09

Ff09 H9

Not available.

off

Does not detect a human presence.

F10 Human

sensor

Ff10

on

Detects a human presence.

F11 Automatic

Ff11 off

Does not output the results
automatically.

Advertising
This manual is related to the following products: