17

8. COMMUNICATION SPECIFICATIONS

The TM-2655/TM-2655P is equipped with two RS-232C channels. Various settings for each channel are
available in the function setting mode. Refer to “7-2 Description of the Functions” for details.

8-1 Channel 1 : Miniature DIN 8-pin

CAUTION

z The personal computer and medical equipment connected to the device are not allowed to

be in the patient area.

z The personal computer used must conform to IEC60950

Specifications

Table 6 Channel 1 specifications

Standard

Conforms to EIA RS-232C

Transmission Asynchronous,

half-duplex

Baud rate

1200, 2400, 4800, 9600 bps (Can be changed in the function setting mode “F07”.)

Data bits

8 bits

Parity bit

None

Stop bits

1 bit

Code ASCII

Equipment available for connection

Channel 1:

Scales, and automatic weight and height scales manufactured by A&D

Channel 2:

Personal computer

Pin assignment

Pin No.

Signal name Description

1 TXD

Send

data

2 RXD

Receive

data

3

RTS

Ready to send

4 -

Not

connected

5

CTS

Clear to send

6 GND

Signal

ground

7 -

Not

connected

8 -

Not

connected

Fig.14 Pin assignment

Cable connection between the device and a personal computer

TM-2655/TM-2655P Personal computer

Note

Do not use pins 4, 7, and 8.
They are used by the device.