Boonton RF Power Sensor/Voltage Probe Security Procedures

848000 Rev B, 03/15/2011

Boonton RF Power Sensor/Voltage Probe Security Procedures

This discussion covers the following Boonton Electronics models: All peak and CW RF Power Sensors
and all RF Voltage Probes. All current Boonton power sensors and voltage probes contain non-volatile
memory for the purpose of storing sensor and calibration data.

1. Size/type of memory

a. Maximum Non-volatile EEPROM, 16384-bit (2Kx8)

2. Location of memory

a. Peak power sensors: Inside sensor housing

b. CW power sensors: "Smart Adapter" or cable plug at instrument end of sensor cable

c. RF voltage probes: "Smart Adapter" or cable plug at instrument end of sensor cable

3. Contents of memory

a. Sensor/probe characteristics (type, operating parameters)

b. Manufacturing and calibration dates

c. Model and serial number

d. Linearity calibration data

e. Frequency calibration data

f.

Temperature calibration data

g. Factory or calibration lab identification string

4. Memory read access

EEPROM information is uploaded to the power meter or voltmeter after the sensor/probe is
connected to the instrument. Certain data may be viewed on display, and all data may be read via
instrument's remote interface using published commands.

5. Memory write access

There is no user accessible EEPROM memory in Boonton Power sensors, or RF probes. The
WTG (Boonton) Factory calibration centers can write (3a) through (3g) data to the EEPROM via
special software during the calibration, but the software and write procedures are not available to
anyone other than authorized factory personnel.

6. Sanitization procedure

There is no user accessible EEPROM memory and as a result there is no memory sanitization
procedure for Power Sensors, or RF probes.