Sper Scientific 850034K AquaShock® Water Purity Kit User Manual

Page 16

Advertising
background image

17

View Calibration Points

1. Press

RANGE to choose the scale you wish to view.

2. Press

SETUP to enter the Setup Mode.

3. Press ↑ or ↓ to select the View Cal Data screen (see Fig. A).

4. Press

SETUP to enter the View Cal Data Mode. The date

and time is a cyclic display (the Conductivity scale will

display as in Fig. B and Fig. C. The TDS scale will display

as in Fig. D and Fig. E). If there are no calibration points

set the display will show “- - - -” (see Fig. F).

5. Press ↑ to view the previous Cal point. Repeat to

view all previous Cal points.

Press ↓ to view the next Cal point. Repeat to view

additional Cal points.

6. Press

SETUP or EXIT to return to Normal Mode.

Fig. A

Fig. B

Fig. C

Fig. D

Fig. E

Fig. F