Sper Scientific 840085 Manometer - 2 PSI User Manual

Page 2

Advertising
background image

2

CONTENTS


Title Page

1. Introduction ..........................................................................

2


2. Panel Description .................................................................

3


3. Instructions

3-A. General

Instructions ..................................................

3-4

3-B. Zero Setting ..............................................................

4

3-C. Calibration .................................................................

4-5

3-D. Auto Power Off

..........................................................

5

3-E. Battery Replacement ................................................

5

3-F. USB

Output ...............................................................

5

3-G. Optional Accessories ................................................

5

3-H. Screen Examples ......................................................

6

3-I. Troubleshooting

........................................................

6

4. Specifications

4-A. Model Range and Resolution ..................................

7

4-B. General Specifications .............................................

7

4-C. Calibration Reference Point ......................................

7

5. Warranty

.............................................................................

8




1. INTRODUCTION

Ideal for measuring low differential pressure in clean rooms, test and
balance, medical equipment, HVAC, pneumatic systems and computer
peripherals. Model 840085 reads gauge pressure and differential pressure
(Imperial and Metric) in 5 units of measure: bar, psi, inH

2

O, mmH

2

O, and

mbar. All other models have 11 units of measure: bar, mmHg, ozin

2

, kgcm

2

,

psi, inH

2

O, kPa, ftH

2

O, inHg, cmH

2

O, and mbar. Features include a large

backlight display, max-min, hold, and auto power off functions. The USB
port enables communication with a computer. High accuracy of ±0.3% fs at
25ºC, repeatability: ±0.2% (max ±0.5%) FSO, and combined linearity and
hysteresis: ±0.2 (Max ±1.0%) FSO.

CAUTION: During the Zero Setting (page 4). Make sure the tube is
disconnected and do not apply pressure to the connector. Doing so can
cause permanent damage to the unit, which is not covered by the warranty.