INTRODUCTION

This simple and easy to use waterproof
pH Pen measures from 2~12 pH.

DISPLAY / FACEPLATE

CALIBRATION

 Calibrate routinely before use, an-

ytime the readings seem erratic, and
whenever the batteries are replaced.

 Remove the protective cap.

 Clean the electrode with de-ionized/

distilled water. Note: Crystals may
form in the cap and these can be
soaked or rinsed away. Do not touch
or rub the glass bulb.

 Before first use or after extended

storage, soak the electrode in
distilled water for 10~30 minutes.

 Press >1 sec to turn on the unit.

 For 1-point calibration, dip/stir the

electrode into pH 7 buffer solution
and press the button. If the
probe is damaged or cannot detect
the buffer, the unit automatically exits
calibration mode after 10 seconds.

 During calibration, "CA" flashes on

the display.

 Then, "CAL" and the pH value are

shown.

 "SA" (save) is displayed as the

calibration point is saved.

 "En" (end) displays when the

calibration ends.

 For 2 or 3 point calibration, start with

pH 7, then repeat the above steps for
pH 4 and/or pH 10.

 For highest accuracy, calibrate at the

sample temperature.

 Repeat the above procedures until

readings stabilize.

 NOTE: If the buffer value is not

standard (for example 4.2), press ▲
to adjust the displayed value.
"SA" (save) is displayed, followed by
"En" (end).

MEASUREMENT

 Calibrate the pen as needed.

 Remove the protective cap and clean

the electrode with de-ionized/distilled
water. Do not touch the glass bulb.

 Press >1 sec to turn on the unit.

 Dip/stir the electrode into the sample.

 The pH value is displayed.

 "- -

-

" is displayed when the pH

sample is out of range.

 During measurement, press ▲ to

freeze the reading on the LCD.
"Hold" is displayed. Press ▲ again to
return to normal measurement.

 The pH Pen automatically shuts off

after 20 minutes of inactivity. To
disable this feature, start with the unit
off, then simultaneously press the
and ▲ buttons. Release both buttons
when "n" is displayed. The unit
returns to measurement mode.

 After use, turn off the unit and rinse

the electrode.

 Protect the electrode's lifespan by

adding a few drops of KCL (potassium
chloride) solution into the protective
cap before storage.

 Always keep the protective cap on

when the unit is not in use.

Rev. 3/29/2011

Pull to

Remove the

Protective Cap

Electrode

FIRST USE: If the
unit fails to turn on,
check that the
batteries are working
and properly
installed.

HoldCal

  7.0 

ATC

87.

5

C
F

pH Value

Symbol will
flash when
battery is low

Hold
button

Calibration
button

On/Off
button

HLD

CAL