Posey Nurse Call Adapter Cable Sets User Manual

Posey, Nurse call adapter cable set instructions, Nurse call adapter cable set

Advertising
background image

M6253 REV C 100313

© 2013 Posey Company. All rights reserved.

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014



Posey

®

Nurse Call Adapter Cable Set

Instructions

FALL MANAGEMENT

Nurse Call Adapter Cable Set

You can connect the following Posey alarms: Sitter Elite

(8345), KeepSafe Deluxe (8374, 8374NP) and Sitter Select

(8361) to your facility’s nurse call system to communicate

with the nursing station in one of two ways:
1. The 8282 will connect the alarm to a nurse call system if

the wall outlet has a ¼” jack and you want to connect

the alarm only.

2. If the wall outlet is not a ¼” jack or if you need a nurse

call button in conjunction with the alarm, you will then

require the Nurse Call Adapter Cable Set (adapter and

nurse call cable). We offer the following set of part

numbers that include both adapter and nurse call cable

for use in tandem. (See list above.)

To Connect the Nurse Call Cable Only

Insert one end of cable into aperture labeled “Nurse Call”

jack on the appropriate side of alarm (8345 right, 8374 left,

8361 left).

Insert other end of cable into wall jack of nurse call panel.

Connecting the Adapter to Nurse Call Cable

If you require an adapter to connect to the nurse call cable

to communicate with the nursing station:

Insert one end of the nurse call cable into one socket of

the adapter. Then place the nurse call cable system

connector into the wall outlet. Plug your existing call cord

into the other socket of the adapter.

Alarm with Nurse Call Cable

REF

8235NCA

REF

8235NCC

REF

8235NCD

REF

8235NCF

REF

8235NCG

REF

8235NCHH

REF

8235NCJ

REF

8235NCJJ

REF

8235NCK

REF

8235NCL

REF

8235NCM

REF

8325NCMM

REF

8235NCNN

REF

8235NCP

REF

8235NCPP

REF

8235NCR

REF

8235NCRR

REF

8235NCS

REF

8235NCVV

REF

8235NCWW

REF

8235NCXX

When using a nurse call cable, ensure the

nurse call cable is plugged into both the alarm and wall

jack before leaving the patient unattended. Verify that

an alert is received at the nursing station if the cable is

unplugged from the wall jack.

NOTE: There will be no alert at the nursing station or at

the bedside if the nurse call cable is unplugged from the

alarm.

• Do not stretch or strain cable to avoid possible

damage and possible malfunction.

• Do not attach cable to moving parts of the bed or

chair that will cause strain or damage if the bed or

chair is repositioned.

• Always position the cable so that moving parts (side

rails, wheels, etc.) will not cause strain or damage the

cable.

• Do not run over the cable with carts or equipment.
• Do not wrap the cable tightly during storage.
• Always remove the cable by pulling on the jack. Do

not pull on the cable.

• Always secure the cable out of the way so that it will

not be a tripping hazard.

• Always test alarm and nurse call function prior to

leaving the patient unattended. Activate the alarm

(remove pressure from sensor, unfasten chair belt

sensor) and make sure the nurse call light for the

proper bed and room activate in the hall and at the

nurses’ station.

• Do Not use the alarm sensor if they don’t activate

each time weight is removed from the sensor or the

chair belt is unfastened. Replace alarm and sensor

with working units and retest before leaving patient

unattended.

8235NCS

Advertising
This manual is related to the following products: