Posey Cufflator™ Endotracheal Tube Inflator and Manometer User Manual

Posey, Instructions for use, Cufflator

Advertising
background image

Instructions for Use:

1. The Posey Cufflator should only be used on tracheal tubes with

high-volume, low-pressure cuffs.

NOTE: The Posey Cufflator is

designed only for use with air-filled cuffs. Use with saline-filled

cuffs will cause damage to the unit and void the product

warranty.

2. Before use, the Control Inflator needs to be checked as follows:

a. Close connecting piece with the finger (Fig. 1).

b. Inflate with inflation bulb to 120 cm H

2

0; value must be

constant for 2-3 seconds. If the pressure drops, the device

needs repair by the Posey Company.

c. Inspect the unit and check the cuff for leaks prior to use.

Prior to intubation or extubation, withdraw all the air

from the cuff with a syringe and close the inflation line.

3. To properly seal:

a. Connect the patient to the ventilator.

b. Connect the Posey Cufflator to the cuff inflation line

and inflate the cuff to a pressure in the range of 60-90

cm H

2

0. This will ensure that the cuff is in close contact

with the tracheal wall.

c. Immediately release air by pressing the red button (Fig. 2)

until the lowest safe cuff pressure level is reached.

d. Intra-cuff pressure should be maintained at a minimum of

20-25 cm H

2

0 to reduce the occurrence of microaspiration

1

and a maximum of 34 cm H

2

0 to decrease the incidence of

mucosal ischemia and subsequent stenosis

2

.

4. The extension tube may be used if constant monitoring of

intra-cuff pressure is desired. Connect the extension tube to the

cuff inflation line and to the Posey Cufflator. Use the hook on the

back of the Posey Cufflator to hang it on the headboard of the

bed. The Posey Cufflator will now monitor the intra-cuff pressure

continuously and can be inflated or deflated as required.

5. The accuracy of the Posey Cufflator may be verified by

connecting it to a mercury sphygmomanometer. Remember

that the Posey Cufflator is calib rated in cm of water pressure

(H

2

0). The conversion rate is 1 mm of mercury (Hg) equals 1.36

cm water (H

2

0). For example, the Posey Cufflator will read 20

cm H

2

0 when the mercury sphygmomanometer reads 14.7

mmHg.

6. The accuracy of measurements is + 2 cm H

2

0 for the whole

range. The Cufflator should be calibrated annually, or if

measurements fall outside of this range, or if the Cufflator

needle does not indicate a reading of zero when nothing is

connected, or if the unit is ever dropped.

DO NOT use with saline!

RX ONLY. Federal law (USA) restricts this device to sale by or on

order of a physician.

Cleaning Instructions:

Wipe the surface thoroughly with an alcohol based disinfectant.

DO NOT submerge the Posey Cufflator and DO NOT autoclave. The

Posey Cufflator face should be cleaned with a glass cleaner only.

Storage:

• This device is designed for use in normal indoor environments.

• This device may be stored in ambient warehouse temperatures

at normal humidity levels. Avoid excess moisture or high

humidity that may damage product materials.

Disposal:

Dispose of per facility policy for BIOHAZARDOUS

material. Be sure to follow all laws that apply.

Service/Repair

NEVER open the Posey Cufflator body. If the Posey Cufflator

body is opened, any damages that result will not be covered

under warranty. Return all Cufflators to the Posey Company for

repair (please call first for a return authorization: 1.800.447.6739

or 1.626.443.3143).

1

Elpern EH. Pulmonary aspiration in hospitalized adults. Nutrition in Clinical Practice 1997: 12(1):5-13.

2

Hill BB, Zweng TN, Maley RH, et. al. Percutaneous dilational tracheostomy: Report of 356 cases. Journal of Trauma 1996; 40:238-245.

For additional information see Stauffer, J.L. Complications of Endotracheal Intubation and Tracheotomy. Respiratory Care 1999; 44(7) 828-844.

Cufflator

REF

8199 Cufflator, standard size, complete with extension

tube and instrument case

8199

Fig. 1

Fig. 2

© 2012 Posey Company. All rights reserved.



M1410 REV D 052912

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

Rx ONLY

Posey

®

Cufflator

RESPIRATORY THERAPY

Advertising