Vaisala Calibration of Digital Transmitters with HMI41 User Manual

Page 34

Advertising
background image

User's Guide ______________________________________________________________________

32 __________________________________________________________________ M210777EN-B

RH

Insert the transmitter probe to the lower
humidity.

RH

%

Let the readings stabilize.

MODE

RH

set

%

Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).

V (number up) or
W (number down)
ENTER

RH

Insert the transmitter probe to the higher
humidity.

RH

%

Let the readings stabilize.

MODE

RH

%

set

Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).

▲ (number up) or
▼ (number down)
ENTER to conclude

RH

Calibration has been successful.

ON/OFF
(disconnect the cable)

RH

Calibration has not been successful.
Perform the calibration again.