Calibrate Humidity Measurement

Use the CRH command to perform a one-point or two-point correction to

the humidity measurement of the module. This command changes the

offset and gain of the measured capacitance, and therefore affects

humidity measurement over the entire temperature range. Take care

when making this adjustment, and ensure you make a note of the existing

values before adjusting them.

CRH<cr>

When performing a one-point calibration, you need to place the probe in

a single humidity reference with RH < 50%. Run the command and enter

the exact RH of the reference after the measurement has stabilized. Exit

the command by pressing the ESC key before proceeding to the second

point (see example 1 below).

For two-point calibration, the first point requires a RH < 50% humidity

reference, the second point must be RH > 50%. There must also be at

least 30 percentage points of difference between the references.

Example: one-point calibration

>crh

RH: 11.29915720 Ref1 ? 11.3

Press any key when ready

>

Example: two-point calibration

>crh

RH: 11.29915720 Ref1 ? 11.3

Press any key when ready

RH: 75.04306440 Ref2 ? 75

OK

>