Diagnostic commands

The PTU200 series digital transmitters respond to a set of diagnostic
commands, which are useful for analyzing the basic settings and
general operation of the transmitters. The diagnostic commands are
listed in

T

ABLE

7-2

Diagnostic commands

Function

Command

basic information on settings

?

software version

VERS

serial number

SNUM

outputting error messages

ERRS

? Basic information on the transmitter settings

? <cr>

The ? command lists the basic information of the transmitter.

>?
Software version

PTU200 / 1.02

Serial number

T5030004

Configuration

12

Linear adjustments

ON

Multipoint adjustments

ON

Calibration date

1999-03-12

Baud Parity Data Stop Dpx 9600

E 7 1 F

Echo

ON

Sending mode

STOP

Address

0

Output interval

3 s

Output format

4.1 P " " UUU " " 2.1 RH " " UUU #r #n

Error output format
SCOM format
Pressure unit

hPa

Temperature unit

'C

Transducer 1:

TYPE: PMT

Transducer 2:

TYPE: VMT

Transducer 3:

TYPE: NONE

>

NOTE

Use this command only for one transmitter at a time. Any PTU200
series transmitter will always respond to ?-command whatever its
settings are, provided that the serial interface settings are correct.