Vaisala PTB200 User Manual

Page 25

Advertising
background image

PTB 200

24 (28)

Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

Example of an offset and gain adjustment:

>LP<cr>
Ref1 ? (set first pressure level and enter value
here)<cr>
P : (the current non-corrected value
appears here)
Ref2 ? (set second pressure level and enter
value here)<cr>
P : (the current non-corrected value
appears here)

Use ESC to abort without executing a command.

Return the MEMORY jumper to the EEPROM write disable position.

4.2

Multipoint pressure adjustment

MPC OFF Multipoint correction off
MPCI

Multipoint correction input

MPC ON Multipoint correction on

MPC OFF<cr>
MPCI<cr>
MPC ON<cr>

The MPC/MPCI commands are used in multipoint pressure adjustment
at up to eight pressure levels.

If a new fine adjustment is needed, the previous multipoint corrections
must first be omitted with the MPC OFF command. Writing down the
previous multipoint corrections is recommended as they will be lost
when MPCI command is used. Precalibrate the barometer to find out
the new multipoint corrections; the corrections are then entered using
the MPCI command. The MPC ON command is used to activate the
new corrections.

Change the MEMORY jumper to the EEPROM write enable position.

Example of the multipoint pressure correction procedure:

>LIP<cr>
P offset 0 * 0.001 ? 0<cr>
P gain 0 * 0.00001 ? 0<cr>
>MPC OFF <cr>
Multip. corr. : OFF