.ID Setting a Identification Code

.ID.xxx<cr>

xxx =

identification code (max. 15 characters)

If there are several barometers on same line, it is necessary to use the
ID -code in order to direct the command for a specific instrument.

NOTE

As default, the ID code is 0. When several barometers will be used on
same line, the codes have to be set before hand separately for each
instrument.

The barometer handles the ID-code as characters i.e. 0001, 01 and 1
are different codes

Setting Pressure Limits

If desired a lower and a higher pressure limits can be given to the
barometer. The limits can be used to define e.g. the calibrated pressure
range to a host system.

.PMIN.xxxx<cr>

xxxx =

lower pressure limit (0 ... 15000 hPa)

If current pressure is under the limit, the barometer outputs ****.** in
stead of the pressure reading.

.PMAX.xxxx<cr>

xxxx =

higher pressure limit

If current pressure is over the limit, the barometer outputs ****.** in
stead of the pressure reading

NOTE

To activate the new settings, please remember to reset the barometer
with the command .

RESET

<cr>