Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 123

Table 27

Relay State Examples

Criterion for

Relay Activation

Transmitter

Operation Status

NC-C

Connected

NO-C

Connected

Not chosen

Power off

aw below 0.5*

Measured aw
above 0.5

Measured aw
below 0.5

aw above 0.5*

Measured aw
above 0.5

Measured aw
below 0.5

Fault status

Measurement OK

Error active

Online status

Measurement live

Error active

* Hysteresis has an effect when relay switching is based on live measurement.
See section Hysteresis on page 121.

Enabling/Disabling Relays

You can deactivate the relay outputs for example for service purposes of
your system.

Setting Relay Outputs

NOTE

When you have only one relay module installed, its relays are called
“relay 1” and “relay 2”.

When you have two relay modules, the relays of the module connected
to slot MODULE 1 are called “relay 1” and “relay 2” and relays
connected to slot MODULE 2 are called “relay 3” and “relay 4”.