Communication Format

Data is transferred from the meter through a serial communication channel.

In serial communications, the voltage is switched between a high and low level

at a predetermined rate (baud rate) using ASCII encoding. The receiving device

reads the voltage levels at the same intervals and then translates the switched

levels back to a character.

The voltage level conventions depend on the interface standard. The table

lists the voltage levels for each standard.

Data is transmitted one byte at a time with a variable idle period between

characters (0 to ∞). Each ASCII character is “framed” with a beginning start bit,

an optional error detection parity bit and one or more ending stop bits. The data

format and baud rate must match that of other equipment in order for

communication to take place. The figures list the data formats employed by

the meter.

Start bit and Data bits

Data transmission always begins with the start bit. The start bit signals the

receiving device to prepare for reception of data. One bit period later, the least

significant bit of the ASCII encoded character is transmitted, followed by the

remaining data bits. The receiving device then reads each bit position as they are

transmitted. Since the sending and receiving devices operate at the same

transmission speed (baud rate), the data is read without timing errors.

Parity bit

After the data bits, the parity bit is sent. The transmitter sets the parity bit to

a zero or a one, so that the total number of ones contained in the transmission

(including the parity bit) is either even or odd. This bit is used by the receiver

to detect errors that may occur to an odd number of bits in the transmission.

However, a single parity bit cannot detect errors that may occur to an even

number of bits. Given this limitation, the parity bit is often ignored by the

receiving device. The meter ignores the parity bit of incoming data and sets the

parity bit to odd, even or none (mark parity) for outgoing data.

Stop bit

The last character transmitted is the stop bit. The stop bit provides a single bit

period pause to allow the receiver to prepare to re-synchronize to the start of a

new transmission (start bit of next byte). The receiver then continuously looks

for the occurrence of the start bit.

IDLE 0 b

0

1

b

2

b b

3

b

4

b

5

6

b

1 IDLE

IDLE

IDLE 0

b

b

0

b

1

3

2

b b

4

b

5

b

6

P 1

IDLE

IDLE 0

b

b b

0

1

b

b

3

2

4

b

b

5

6

1 1

(7 data, no parity, 2 stop)

(7 data, parity, 1 stop)

(8 data, no parity, 1 stop)

Start bit

Stop bit

Note: b - b is ASCII data.

0

7

7

b

Character Frame Figure

LOGIC

RS232*

RS485*

INTERFACE STATE

1

TXD,RXD; -3 to -15 V

a-b < -200 mV

mark (idle)

0

TXD,RXD; +3 to +15 V

a-b > +200 mV

space (active)

* Voltage levels at the Receiver

d-LEV

Display

Intensity Level

PAR

9-FCS

Pro

Factory

Service Code

COdE

6.9 Module 9 - F

actory

s

ervIce

o

peratIons

(



)

PARAMETER MENU

Enter the desired Display Intensity Level (0-15) by

using the arrow keys. The display will actively dim or

brighten as the levels are changed. This parameter also

appears in Quick Programming Mode when enabled.

DISPLAY INTENSITY LEVEL



The meter has been fully calibrated at the factory.

Scaling to convert the input signal to a desired display

value is performed in Module 1. If the meter appears to be

indicating incorrectly or inaccurately, refer to

Troubleshooting before attempting to calibrate the meter.

When recalibration is required (generally every 2 years), it should only be

performed by qualified technicians using appropriate equipment. Calibration

does not change any user programmed parameters. However, it may affect the

accuracy of the input signal values previously stored using the Apply (



)

Scaling Style.

Calibration may be aborted by disconnecting power to the meter before

exiting Module 9. In this case, the existing calibration settings remain in effect.

CALIBRATION









RESTORE FACTORY DEFAULTS

Use the arrow keys to display

 

and press

PAR.

The meter will display



and then return to

 

.

Press

DSP key to return to Display Mode. This will

overwrite all user settings with the factory settings.

Input Calibration

WARNING: Calibration of this meter requires a signal source with an

accuracy of 0.01% or better and an external meter with an accuracy

of 0.005% or better.

Before starting, connect -SIG (terminal 3) to COMM (terminal 4).

This allows a single ended signal to be used for calibration. Connect the

calibration signal to +SIG (terminal 2) and -SIG (terminal 3). Verify the Input

Range jumper is in the desired position. Allow a 30 minute warm-up period

before calibrating the meter.



and

PAR can be chosen to exit the calibration

mode without any changes taking place. Perform the following procedure:

1. Press the arrow keys to display

 

and press

PAR.

2. Choose the range to be calibrated by using the arrow keys and press

PAR.

3. When the zero range limit appears on the display, apply 0 mV between +SIG

and -SIG.

4. Press

PAR and ---- will appear, wait for next prompt.

5. When the top range limit appears on the display, apply the corresponding

+SIG and -SIG voltage (20 mV or 200 mV).

6. Press

PAR and ---- will appear, on the display for about 10 seconds.

7. When



appears, press

PAR twice to exit programming.

8. Repeat the above procedure for each range to be calibrated or to recalibrate

the same range. It is only necessary to calibrate the input ranges being used.

9. When all desired calibrations are completed, remove -SIG to COMM

connection and external signal source.

10. Restore original configuration and jumper settings.