SPECIFICATIONS

POWER:

AC Power: 40 to 250 VAC, 50/60 Hz, 20 VA

DC Power: 21.6 to 250 VDC, 8 W

Isolation: 2300 Vrms for 1 min. to all inputs and outputs.

INPUT CAPABILITIES:

Current Input Ranges:

± 250 µADC

± 2.5 mADC ± 25 mADC

± 250 mADC

± 2 ADC

Voltage Input Ranges:

± 250 mVDC

± 2.0 VDC

± 10 VDC

± 25 VDC

± 100 VDC

± 200 VDC

Thermocouple Inputs:

Types: T, E, J, K, R, S, B, N, C (W5/W26)

Max Continuous Overvoltage: 30 V

RTD Inputs:

Type: 3 or 4 wire, 2 wire can be compensated for lead wire resistance

Excitation current: 100 ohm range: 136.5 µA ±10%

10 ohm range: 2.05 mA ±10%

Max. continuous overload: 30 VDC

Input Type:

100 ohm Pt alpha = .00385

100 ohm Pt alpha = .00392

120 ohm Nickel alpha = .00672

10 ohm Copper alpha = .00427

Resistance Inputs:

Max. continuous overload: 30 VDC

INPUT RANGE

COMPLIANCE

100 ohm

0.175 V

999 ohm

1.75 V

9999 ohm

17.5 V

EXCITATION POWER: Jumper selectable

Transmitter Power: +18 VDC, ± 5% @ 50 mA max.

Reference Voltage: + 2 VDC, ± 2%

Compliance: 1KW load min (2 mA max)

Temperature Coefficient: 40 ppm/ºC max.

Reference Current: 1.05 mADC, ± 2%

Compliance: 10 KW load max.

USER INPUTS: Two programmable user inputs

Max. Continuous Input: 30 VDC

Isolation To Sensor Input Common: Not isolated.

CUSTOM LINEARIZATION:

Data Point Pairs: Selectable from 2 to 16

Display Range: -1999 to 9999

Decimal Point: 0 to 0.000

ENVIRONMENTAL CONDITIONS:

Operating Temperature Range: 0 to 50 °C

Storage Temperature Range: -40 to 60 °C

Vibration to IEC 68-2-6: Operational 5-150 Hz, 2 g

Shock to IEC 68-2-27: Operational 25 g (10 g relay)

Operating and Storage Humidity: 0 to 85% max. RH non-condensing

Altitude: Up to 2000 meters

CERTIFICATIONS AND COMPLIANCES:

CE Approved

EN 61326-1 Immunity to Industrial Locations

Emission CISPR 11 Class A

IEC/EN 61010-1

RoHS Compliant

UL Listed: File #E179259

Type 4X Indoor Enclosure rating (Face only)

IP65 Enclosure rating (Face only)

IP20 Enclosure rating (Rear of unit)

CONNECTIONS: High compression cage-clamp terminal block

Wire Strip Length: 0.3" (7.5 mm)

Wire Gauge Capacity: One 14 AWG (2.55 mm) solid,

two 18 AWG (1.02 mm) or four 20 AWG (0.61 mm)

CONSTRUCTION: This unit is rated NEMA 4X/IP65 for indoor use only.

IP20 Touch safe. Installation Category II, Pollution Degree 2. One piece

bezel/ case. Flame resistant. Synthetic rubber keypad. Panel gasket and

mounting clip included.

WEIGHT: 8 oz. (226.8 g)

Drawing No. LP0903C
Released 04/14

Tel +1 (717) 767-6511
Fax +1 (717) 764-0839
www.redlion.net

3.80

1.95

0.10

4.14

(96.52)

(49.53)

(2.54)

(105.16)

1.75

(44.45)

3.60

(91.44)

Note: To determine dimensions for

horizontal units, swap height and

width. Recommended minimum

clearance (behind the panel) for

mounting clip installation is:

2.1" (53.4) W x 5.5" (140) H.

DIMENSIONS In inches (mm)

C

US LISTED

UL

R

3RSD

PROCESS CONTROL EQUIPMENT

Advertising