NOTE: For TPUDP pumps, a Remote Start option is

available using the 9-pin female D-connector located
on the rear panel, via pins 1 and 5.


Continuous Pulse Mode

1. Using an appropriate flathead screwdriver, dial the Time

Period (white) knob to the desired position, then dial the
Multiplication (blue) knob to the desired position.

2. To run a preset continuous program of a specified run time

(ON) followed by the same length of OFF time, set the relay
timer MUR3 Mode switch to Di.


Remember that you can also set the pumping speed, which
will control the amount dispensed during the ON period.


NOTE: Any time you change the TPUDP pump’s MUR3 timer

Mode, the new setting will be activated only after

cycling Timer’s On/Off switch. We recommend

therefore that you change timer Mode settings when

the timer switch set to the Off position. This way, when the
timer On/Off switch set back to the On position, the new
settings will be activated.