CS106 Barometric Pressure Sensor

6.2 Accuracy

Total Accuracy***:

±

0.3 mb @ +20 °C

±

0.6 mb @ 0 to +40 °C

±

1 mb @ –20 to +45 °C

±

1.5 mb @ –40 to +60 °C

Linearity*:

±

0.25 mb @ 20 °C

Hysteresis*:

±

0.03 mb @ 20 °C

Repeatability*:

±

0.03 mb @ 20 °C

Calibration Uncertainty**: ±0.15 mb @ 20 °C

Long-Term Stability:

±

0.1 mb per year

* Defined as ±2 standard deviation limits of end-point non-linearity,

hysteresis error, or repeatability error

** Defined as ±2 standard deviation limits of inaccuracy of the working

standard at 1000 mb in comparison to international standards (NIST)

*** Defined as the root sum of the squares (RSS) of end-point non-linearity,

hysteresis error, repeatability error and calibration uncertainty at room

temperature

6.3 General

Dimensions:

9.7 x 6.8 x 2.8 cm (3.8 x 2.7 x 1.1 in)

Weight:

90 g (3.2 oz)

Housing Material:

ABS/PC blend

Supply Voltage:

10 to 30 Vdc

Supply Voltage Control:

When the internal jumper is closed, the CS106

is on continually. When the jumper is open,

the CS106 can be turned on/off with 5 Vdc/

0 Vdc.

Supply Voltage Sensitivity: negligible

Current Consumption:

<4 mA (active); <1 µA (quiescent)

Output Voltage:

0 to 2.5 Vdc

Warm Up Time:

1 s

Pressure Fitting:

barbed fitting for 1/8 in I.D. tubing

Overpressure Limit:

2000 mb