2 step 2 — do a false echo learn command, Step 2 — do a false echo learn command, 4. description of polarization markings labels – Campbell Scientific CS475, CS476, and CS477 Radar Water Level Sensor User Manual

Page 12

Advertising
background image

CS475, CS476, and CS477 Radar Water Level Sensor

CS475

CS476/CS477

1

2

FIGURE 4-1. Polarization markings (see TABLE 4-4 for label

descriptions)

TABLE 4-4. Description of Polarization Markings Labels

Sensor

Description

1 CS475

Polarization marks are designated by the mounting loop
screws.

2 CS476/CS477

Polarization mark is machine-tooled.

4.2 Step 2 — Do a False Echo Learn Command

The 25616, Adjustment/Display Module, or the terminal emulator
in LoggerNet or PC400 can be used to enter SDI-12 commands.
Appendix B.2, Using Terminal Emulator and a Datalogger to
Send Commands
,
describes entering commands using the
terminal emulator.

NOTE

Do a Start False Echo Learn command followed by a Send Data command
(see TABLE 4-4). To start false echo learn, do the aXSFEL+nnn.nnn!
command (where nnn.nnn = the actual distance to the water) followed by the
aD0! (Send Data) command. TABLE 4-5 shows an example of the command
and response.

Any echo occurring 0.5 m (1.6 ft) short of the distance you entered will be
considered noise.

Appendix B describes this command in further detail.

4

Advertising