Appendix B. CR23X Wiring and

Programming

B.1 CR23X Connections

Reconfigure the receiver for 1200 baud using Garmin configuration software

available on Garmin’s website. Refer to TABLE B-1 to connect the GPS16X-

HVS directly to a CR23X datalogger.

TABLE B-1. CR23X Wiring

GPS16X-HVS

CR23X

Function

Red

12V

Power In

Black

Ground

Power Ground

Yellow

Ground

Power Switch

White

Control Port

TXD

Gray

Control Port

PPS

Blue

Ground in operation

Send data to receiver for

configuration

R

x

data

Shield

Ground

Shield

B.2 Programming

Program instruction 15 (P15) is used to read the NMEA $GPGGA string of

time and position data. Each iteration of P15 can either read the numeric fields

or read everything. When reading the numeric fields, such as time, latitude,

longitude and elevation, P15 requires non-numeric delimiters between data

points. The only available format of GPS data with delimiters is the NMEA

0183 format. Program instruction 15 (P15) reads serial data and discards non-

numeric values. All non-numeric values act as delimiters between numbers,

and decimal points can also act as delimiters. P15 can be used to import

everything in the string, character by character, and convert it to the decimal

equivalent. The decimal equivalent method is seldom used, and only when the

general area (hemisphere) is not known.

B.2.1 Program Execution Interval

When the PPS signal is used to trigger the read data function (P15), the

program table execution interval does not matter. Otherwise the timing

between the GPS16X-HVS output and the datalogger read must be considered.

Generally the execution interval can not be less than 2 seconds when the PPS

signal is not used.

B-1