1 temperature sensor, 2 relative humidity sensor – Campbell Scientific HMP60 Temperature and Relative Humidity Probe User Manual

Page 15

Advertising
background image

HMP60 Temperature and Relative Humidity Probe

Probe Body Diameter:

1.2 cm (0.47 in)

Filter: 0.2

μm Teflon membrane

Filter Diameter:

1.2 cm (0.47 in)

Housing

Body Material:

AISI 316 stainless steel

Filter Cap Material:

Chrome-coated ABS plastic

Classification: IP65

Power Consumption:

1 mA typical; 5 mA maximum

Supply Voltage:

5 to 28 Vdc

Settling Time after
power is switched on:
1

second

Output Signal Range:

0 to 1 Vdc

6.1 Temperature Sensor

Sensor: 1000

Ω PRT, DIN 43760B

Temperature
Measurement Range:
-40

° to +60°C

Temperature Accuracy:

±0.6°C (-40° to +60°C)

6.2 Relative Humidity Sensor

Sensor: INTERCAP

®

Relative Humidity
Measurement Range:

0 to 100% non-condensing

Accuracy at 0

° to +40°C:

±3% RH (0 to 90% Relative Humidity)

±5% RH (90 to 100% Relative Humidity)

Accuracy at -40° to 0°C
and +40° to +60°C:

±5% RH (0 to 90% Relative Humidity)

±7% RH (90 to 100% Relative Humidity)

The black outer jacket of the cable is Santoprene

®

rubber.

This compound was chosen for its resistance to
temperature extremes, moisture, and UV degradation.
However, this jacket will support combustion in air. It is
rated as slow burning when tested according to U.L. 94
H.B. and will pass FMVSS302. Local fire codes may
preclude its use inside buildings.

CAUTION

9

Advertising