PWC100 Present Weather Sensor Calibrator

Step 4: Position 2: Left

• Move the calibrator into the left hand position by unscrewing the calibrator

mounting screws slightly and sliding the calibrator fully to the left. Then
retighten the screws once in position.

• Visually check that the calibrator is located correctly and securely.

• LED L should now be flashing slowly and all other LEDs should be off.

• Once you are ready, press START.

• The test is now automated and the disk should start rotating. LED L will

stay illuminated until this stage of the test is complete.

• Wait for the calibration disk to stop rotating. This test will take

approximately two and a half minutes.

• Upon successful completion of the test, LED R will be flashing slowly and

LED L will be turned off.