Appendix B. TDR100 Response Protocol

B-2

Error Response

|!|NN|

! ASCII ‘!’ to indicate an Error Packet

NN

Error Numbers in ASCII

‘01’

Bad

Checksum

‘02’

Illegal Cmd Format, Not Defined

‘03’

No Valid Letters Or Numbers

‘04’

Could Not be Parsed

‘05’

Command

Not

Identified

‘06’

Command

Not

Recognized

`07’

Calibration

Unsuccessful

‘08’

Extra

Period

(terminal

mode)

‘09’

No Reference Cable Length

‘10’

Value Out of Range

‘11’

Timeout - Cable Short Not Found

‘12’

Timeout Waiting for Data

‘13’

Exponent

Not

Defined

‘14’

No Command Defined for Output

‘15’

Bad

Data

‘16’

Bad

Moisture

Calculation

‘17’

Could Not Detect Liquid Level

‘18’

Incorrect Mux Address or Channel

‘19’

Unable to Locate Pulse

‘20’

Could Not Measure Baseline

‘21’

Couldn’t Measure Top of Pulse

‘22’

Unknown Internal Error

‘69’

Command

Decode

Error

‘70’

Unknown

Error

‘71’

Device Write not accepted

‘72’

Unknown

Error

‘73’

IOPOLL:

Timeout

’74’

WRITE: Data not written

‘75’

WRITE: Address not accepted

‘76’

WRITE: Write not accepted

‘77’

Unknown

Error

‘78’

READ: Read not accepted

’79’

READ: Address not accepted

‘80’

Unknown

Error

Advertising