TDR100

TABLE 6-1. Reflection waveform array header elements.

Description of Array Header Elements

1 averaging

2 propagation

velocity

3

number of data points

4 cable

length

5 window

length

6

probe rod length

7 probe

offset

8 multiplier

9 offset

6.3.12 Parameter 12: Multiplier

Multiplication factor applied to the value stored in the input location specified
in parameter 11. Set to 1 when parameter 2 is 1 or 2. The multiplier can be
used for the probe constant value when measuring electrical conductivity.

6.3.13 Parameter 13: Offset

Offset value added to the value stored in input location specified in parameter
11. Set to 0 when parameter 2 is 1 or 2.

6.3.13.1 Probe Constant for Electrical Conductivity Measurement

The section, TDR Principles, presents the equation used for soil bulk electrical
conductivity. For convenience, it is repeated here.

σ

ρ
ρ

=


+

K

Z

c

1

p

1

A portion of the datalogger algorithm evaluates the reflected waveform to
determine the applied and reflected voltages. From this, the reflection
coefficient,

ρ, can be calculated. The term, Z

c

, is the reflectometer impedance

and is a constant 50 ohms.

The electrical conductivity routine in the datalogger returns the value

1 1

1

Z

c

− ρ

+ ρ

which must be multiplied by the probe constant, K

p

, to get an electrical

conductivity value.