11. TROUBLE DIAGNOSIS

SYMPTOM

DIAGNOSIS

REMEDY

Final result is always
zero or near zero

Temperature too low

Raise temperature,

see pages 25 & 68

No heat

See No Heat symptom later in
table

Sample too small

Increase sample size,

see pages 29 & 66

Final results vary too
much

Test temperature incorrect

Determine
temperature/ending criteria to
use or call AZI at
800-528-7411 or
602-470-1414 for assistance

Ending criteria incorrect

Sample size too small

Increase sample size,

see pages 29 & 66

Samples not uniform

Grind sample,

see page 65

Sample particles too large

Blank display or very
hard to see

Contrast controls not set

Adjust contrast, see page 16

No power

Check power cord
connections, see page 14

Microprocessor failure

Turn unit OFF and back ON.
Call AZI at 800-528-7411 or
602-470-1414 for assistance

Backlight connector
disconnected

Call AZI at 800-528-7411 or
602-470-1414 for assistance

Heat will not turn off

Circuit board failure

Call AZI at 800-528-7411 or
602-470-1414 for assistance