VAPORIZER INSTALLATION


The Plasma Freezers require the installation of the condensate vaporizer, which is shipped inside
the unit compartment. The vaporizer should be mounted to the cabinet bottom. See View 1. In lieu
of the vaporizer, these units can be plumbed to a floor drain following local plumbing codes.

Note: Disconnect power
prior to installation.

1.

Connect the power cord
to the vaporizer and
place it in the vaporizer
bracket.


2.

Attach the vaporizer to
the unit using the four
screws as shown. Make
sure the plastic drain
tube is in the vaporizer
pan.









View 1



LOCATION


The refrigeration system located at the top of the cabinet requires free air access for proper
operation. Allow a minimum four-inch clearance on the top, rear, and sides of the cabinet. The
cabinet should also be leveled when it is placed in its permanent location.


ELECTRICAL

Check the proposed outlet to be used to insure that the voltage, phase, and current carrying
capacity of the circuit from the electrical panel correspond to the requirements of the cabinet.
NEVER use an extension cord to wire any unit. On permanently connected units, those not
furnished with a plug-in service cord, all inter-wiring between the electrical panel and the unit must
be done in accordance with the National Electric Code and all state and local codes. See serial tag
for electrical data.

2-10VDC temperature retransmit is standard on some models, optional on others. See the
following pages for details and also see the controller programming manual for configuration details.
An optional 4-20mA temperature retransmit is also available.