Operation – Nor-Lake Large Capacity CO2 Incubator User Manual

Page 4

Advertising
background image

Figure A

CO

2

Operation

The Large Capacity Incubator is offered with a CO

2

dispensing and measurement system. The CO

2

system is microprocessor based and controlled by a sensor based on the NDIR Single-Beam Dual-
Wavelength principle. The sensor (included with CO

2

option) measures from 0-20% CO

2

density

and provides a reliable and stable reference measurement over time. Water vapor, dust, debris and
most chemicals do not affect measurement accuracy; therefore, the sensor may be used in a harsh
and humid environment.

The CO

2

settings are controlled by the controller located on the façade of the unit. Please see the

default and adjustable settings in the separate instructions, part number 136349.

Installation Requirements – The end user must supply two CO

2

tanks and two-two stage gas

regulators for the CO

2

input. See Figure B for hook up points. The input gas pressure must be

above 4 psi as the CO

2

tank switch over valve is factory set to recognize an empty tank when the

pressure drops to 4 psi. The maximum recommended input pressure should not exceed 25
psi.

Connecting a CO

2

Supply

High concentrations of carbon dioxide can cause
asphyxiation. The use of CO

2

monitors and alarms

is recommended for areas where CO

2

can collect.

!

4

11/10 Rev. B 143314