Nor-Lake Undercounter BOD Refrigerator Incubator User Manual

Page 12

Advertising
background image

Page 12

04/14 Rev. B 151749

Pinout

Parameter List

Register Variable Description

Range

***Analog***

1

AIRT

Air Temperature

Read Only

2

PROT

Product Temperature

Read Only

4

ASET

Air Temperature Set Point

2.0 to 50.0ºC

5

HITA

High Air Temperature Alarm Set Point

-50.0 to 75.0ºC

6

LOTA

Low Air Temperature Alarm Set Point

-50.0 to 75.0ºC

7

HITP

High Product Temperature Alarm Set
Point

-50.0 to 75.0ºC

8

LOTP

Low Product Temperature Alarm Set
Point

-50.0 to 75.0ºC

***Integer***

129

ATAD

Air Temperature Alarm Delay

0 to 999
seconds

130

PTAD

Product Temperature Alarm Delay

0 to 999
seconds

133

BURB

Buzzer Ring Back Time

0 to 20 minutes

134

BUTO

Buzzer Tone

0 = Constant
1 = 1 sec on 2

sec off

2 = 2 sec on 5

sec off

163

---

System Status

1 = On
2 = Off: Switch
3 = Off: Alarm

***Digital***

1

---

Air Temperature Sensor Fail Alarm

0 = Off; 1 = On

2

---

Air Temperature High Alarm

0 = Off; 1 = On

3

---

Air Temperature Low Alarm

0 = Off; 1 = On

4

---

Product Temperature Fail Alarm

0 = Off; 1 = On

5

---

Product Temperature High Alarm

0 = Off; 1 = On

6

---

Product Temperature Low Alarm

0 = Off; 1 = On

10

---

Door Ajar Alarm

0 = Off; 1 = On

21

---

Power Reset

0 = Off; 1 = On

Pin

Description

1

GND

2

RX+/TX+

3

RX-/TX-