Pulsafeeder PULSAtron Series HV EN User Manual

Page 11

Advertising
background image

11

For the 4-20 mA input to have any effect on the pump output rate, the AUTO/OFF/MANUAL switch must be in the AUTO
position.

The 4-

20 mA input signal affects the pump’s outputs as per the graph below:

The signal cord polarity is:

Black = Common

White = Positive

Wrong polarity can result in excess flow.

Signal input impedance is 124 ohms.

Remove cap from pump socket labeled 4-20 mA, use polarized cord supplied with pump to connect control circuit to
pump. Plug cord into pump socket labeled 4-20 mA.

6.0 MAINTENANCE

Before performing any maintenance or repairs on chemical metering pumps, be sure to disconnect all

electrical connections, insure that all pressure valves are shut off and pressure in the pump and lines has been
bled off.

Always wear protective clothing, gloves and safety glasses when performing any maintenance or repairs on
chemical metering pumps.

6.1 Routine Maintenance

Routinely check the physical operating condition of the pump. Look for the presence of any abnormal noise, excessive
vibration, low flow and pressure output or high temperatures [when running constantly at maximum stroke rate, the pump
housing temperature can be up to 160

˚F (70˚C)].

For optimum performance, cartridge valves should be changed every 6-12 months. Depending on the application, more
frequent changes may be required. Actual operating experience is the best guide in this situation.

Repeated short-term deterioration of valve seats and balls usually indicates a need to review the suitability of wetted
materials selected for the application. Contact the supplier for guidance.

Check for leaks around fitting or as a result of deteriorating tubing e.g. when standard white translucent discharge tubing
is exposed to direct sunlight. Take appropriate action to correct leak by tightening fittings or replacing components.

Keep the pump free of dirt and debris as this provides insulation and can lead to excessive pump temperatures.

If the pump has been out of service for a month or longer, clear the pump head valve assemblies by pumping fresh water
for approximately 30 minutes.

If the pump does not operate normally after this “purging run”, replace cartridge valve

assemblies.

6.2 Disassembly And Assembly

Diaphragm Removal

Flush pump head and valve assemblies out by running pump with water or other suitable neutralizing solution. Wash
outside of pump if chemical has dripped on pump. Set stroke length knob of pump to 0% and unplug pump.

Advertising