Myron L 767 User Manual

Page 3

Advertising
background image

SECTION 1

Introduction

1.1

SCOPE

This manual provides the user with the necessary information to
install, operate and maintain the Myron L Company’s 750/760
Series Conductivity Monitors. Sections 1 through 3 provide Monitor
applicational descriptions, mounting, wiring and operational
procedures. Section 4 identifies their primary components and
provides the user with easy-to-use calibration and preventive care
procedures.

Section 5 (Appendix A) provides the 750/760 Series Monitor’s
Replaceable Components Chart.

1.1.1 FUNCTIONAL DESCRIPTIONS

All models have drip/weather-proof housings suitable for panel,
bench or surface mounting. The 750 models are a compact 6.0"
(152mm) x 4.8" (122mm). The 767 model is 6.0" (152mm) x 10.8"
(275mm).

Bright green/red LEDs indicate below/above set point readings.
All models except 756 feature a heavy-duty 10 amp output relay,
operating on either increasing or decreasing readings.

For specific Monitor configurations, reference the following
individual model descriptions.

Model 755
Small controller only. Provides a front panel accessible set point
adjustment knob in place of an analog or digital meter readout
display.

Model 756
Small monitor only. Standard front panel is equipped with a linear
analog meter display only (no relay, LED indicators or set point
check switch).

Model 757
Small analog monitor/controller. Single set point is internal to
discourage unauthorized adjustments. Standard front panel is
equipped with a linear analog meter display and a “SET POINT
CHECK” switch.

Model 758
Small digital monitor/controller. Single set point is internal to
discourage unauthorized adjustments. Standard front panel is
equipped with a digital LCD meter display and a “SET POINT
CHECK” switch.

Model 767
Large analog monitor/controller. Single set point is internal to
discourage unauthorized adjustments. Standard front panel is
equipped with a three (3) range select switch, analog meter display
and a “SET POINT CHECK” switch. Options include three (3) cell
input capability and dual set point control.

1.1.2

APPLICATIONS

1

Reverse Osmosis

2

Process Control

3

Seawater Desalinization

4

Waste Treatment

5

Food Processing

6

Power Plants

7

Laboratories

1.2

SPECIFICATIONS

RANGES:

Refer to Conductivity/ppm Ranges sheet on Page 3 for
Conductivity Ranges.

READOUT:

Model 756: 2

1

/

2

" (63

mm) analog meter

Model 757: 2

1

/

2

" (63

mm) analog meter

Model 758:

1

/

2

" (13

mm) 3

1

/

2

digit LCD

Model 767: 4

1

/

2

" (114

mm) analog meter

ACCURACY:

Model 758:

±

1 % of span

Other Models:

±

2% of span

SENSITIVITY:

0.05% of span

STABILITY:

0.05% of span

REPEATABILITY:

0.1% of span

CALIBRATION CHECK:

Built in

RECORDER OUTPUT:

0-10 VDC @ 5

mA max. (linear); standard on all models

CELL INPUT:

1 (optional 3 Cell Input available on Model 767)

RELAY FUNCTION:

Models 755, 757, 758 and 767:

Single set point control continuously adjustable 0-100% of
span

Indicators:

“ABOVE” (red) and “BELOW” (green) set point LEDs

Contact Rating:

SPDT 10 amp @ 250 VAC, 30 VDC. Relay operates
increasing or decreasing reading (selectable).

Optional on Model 767 only:

Dual set point control, with above specifications

POWER SPECIFICATIONS:

115 VAC

±

15%, 50/60 Hz, 25 mA

220 VAC (User changeable)

DIMENSIONS 750 SERIES:

6.0" (152

mm) H x 4.8" (122 mm) W x 3.8" (96 mm) D

DIMENSIONS 760 SERIES:

6.0" (152

mm) H x 10.8" (275 mm) W x 3.9" (99 mm) D

AMBIENT TEMPERATURE RANGE:

-22

°

F (- 30

°

C) to 140

°

F (60

°

C)

HOUSING CONSTRUCTION:

Fully gasketed heavy-duty ABS for corrosion resistance. Rated
NEMA type 3.

WEIGHT:

750 Series: 2 Ibs. (0.9 kg)

760 Series: 3 Ibs. (1.4 kg)

1

Advertising
This manual is related to the following products:

758, 757, 756, 755