F. RE-SCALE YOUR CTCII CONTROLLER

1. DESCRIPTION

The CTC Series II controller, has been designed for easy field re-

scaleability. To re-scale your 0-5000 µS CTCII to 0-5000 “PPM”

TDS 442™ or visa versa requires the simple slide of a switch.

WARNING: BEFORE STARTING, IF MONITOR/CONTROLLER

IS INSTALLED, ENSURE THE POWER IS OFF. FAILURE TO

DO SO COULD CAUSE DAMAGE TO THE INSTRUMENT, AND

COULD BE HARMFUL OR FATAL TO PERSONNEL. ONLY

QUALIFIED PERSONNEL SHOULD INSTALL OR SERVICE

ELECTRICAL EQUIPMENT.

Physical

NOTE: When opening instrument, remove front panel with care. If

the front panel has already been removed from the enclosure skip

to #3.

1. Using a Phillips Head screwdriver remove the four (4)

screws on the clear cover.

2. Remove clear cover.

3. The front panel is held in place with hook & loop strips

on each side. By carefully lifting at the slot on the top,

the front panel will pull toward you. Do not pull more than

about 8 inches/20CM.

DO NOT pull on control knob(s).

4. Rotate the front panel down so that the back side is

facing you. The Circuit Board is mounted to the back of

the front panel.

5. Locate the switch next to the transformer. See figure

II.F.1.

6. Slide knob toward the desired scale - TDS/ppm or

COND/µM(µS).

7. Recalibrate, see CALIBRATION PROCEDURES, section

V.B.

8. Install (remove) Parts Per Million - TDS label (supplied

inside).

9. Reassemble front panel.

!

CAUTION

- READ FOLLOWING CAREFULLY

Main CB Assembly

Figure II.F.1

SET POINT

HYSTERESIS

RIGHT INC

LEFT DEC

115/230 VAC

SELECT SWITCH

100mA Slow Blow (T.10A)

115 & 230 VAC (V~)

SCALE SELECT

ELECTRONIC

CALIBRATION

MAIN

CALIBRATION

CONTROL

DISPLAY

CALIBRATION

CONTROL

(FACTORY SET)

OPTIONAL

LIMIT

TIMER

CONTROL

MODULE

OPTIONAL

4-20 mA

OUTPUT

MODULE

DIS

CAL

TDS

COND

EC

CTCII

DISPLAY

TRANSFORMER

Yel

CAP

CAP

C4AM

LTM

HYST

Flow Switch

Jumper -

Remove if

flow switch

installed.

SWITCH

FUSE

RELAY

RELAY

BLEED FEED

FEED

PUMP

BLEED

VALVE

POWER

FLOW

SWITCH

SENSOR

0-10VDC

FS

BK WTRD GN NU R- R+

PWR

C GD

PWR

C GD

PWR

C GD

Wickman Fuse 5 Amp

(#3741500041 or equiv)

115 VAC(V~)

14