Myron L EP-10 User Manual

4&5 range conductivity meters, Description

Advertising
background image

description

This manual describes your Myron L EP series meter,

tells you how to use it, and how to keep it working

accurately for many years.

Myron L EP & EP10 Meters are compact, multi-range

instruments which operate on the principle of electrical

conductivity. They each contain a built-in cell cup and will

quickly determine the conductivity of almost any solution.

Both have dials calibrated in micromhos (µmhos)

equivalent to the metric microsiemens (µS). Model EP

features a second dial scale calibrated in

megohms/resistivity for testing ultrapure water.

Both are 3.4” x 4.5” x 4.0” (86 x 114 x 102mm) and

weigh less than one pound (0.45 kg). EP series meters are

completely self-contained. The built-in cell is automatically

temperature compensated. They are powered by one 9

volt battery good for at least 2000 tests, or one year shelf

life.
model ranges
ep

0-0.5, 0-5, 0-50, 0-500, 0-5,000 micromhos (µmhos)

2-30 Megohms

ep10 0-10, 0-100, 0-1,000, 0-10,000 micromhos (µmhos)
accuracy: (Both Models) ±2% of full scale

repeatability: (Both Models) ±1%
temperature compensation to 25ºc:
ep

32-120ºF (0-49ºC)

ep10

50-160ºF (10-71ºC)

By using an RE-10 Range Extender (see

aCCESSoRiES), the maximum range of each meter may

be increased ten times.

The pictures in this manual show the major operating

parts of your Myron L EP series meter. Handle your meter

and identify these parts to become familiar with it.

Battery

Indicator

Range Switch

Cell Cup

Button

Upper Electrode

Meter Zero

Adjust

Pointer

Model EP10 shown

4&5 range conductivity meters

oPERaTing inSTRuCTionS

User’s Manual for Models EP

EP-10

Advertising
This manual is related to the following products:

EP