Myron L D-1 User Manual

Single range dialysate meter, Description

Advertising
background image

Single Range Dialysate Meter™

operating instructions

User’s Manual for Model D -1

this manual describes your Myron L D-1 meter, tells

you how to use it, and how to keep it working accurately

for many years.

Your Model D-1 single range Dialysate Meter™ is a

compact instrument which operates on the principle of

electrical conductivity. it will quickly determine if dialy-

sate concentrations are correct. its dial provides an

expanded 11-15 millimhos (millisiemens) scale for ac-

curate readings.

a unique feature of Model D-1 is the green color band

above the dial. this provides an immediate visual con-

firmation of whether or not the dialysate concentration

is within the patient’s “safe zone”. the color band is

adjustable and can be pre-set by the doctor or techni-

cian for each patient. adjustment is made by turning

the screw on the face of the meter. this screw may be

covered with tape or a label to avoid tampering.

the D-1 is 3.4” x 4.5” x 4.0” (85 x 129 x 126mm) and

weighs less than one pound (0.45kg). D-1 meters are

completely self-contained. the built-in cell is automat-

ically temperature compensated from 60º to 110ºF

(16º to 43ºc). they are powered by two 9 volt batteries

good for a least 2000 tests, or one year shelf life. a light

bulb, which glows when the black button is pushed,

serves as a battery reserve indicator.

the pictures in this manual show the major operating

parts of your Myron L D-1 Dialysate Meter. Handle your

D-1 and identify these parts to become familiar with it.

Battery

Indicator

Plunger

Cell Cup

Pointer

Button

Upper

Electrode

Color Band

Adjust

DeScRiption